New TopShim run:

version = topshim 2.0.0
date = 2023-10-31
time = 17:31:39.842
command = topshim  convcomp tuneaxyz

1D SHIMMING

Parameters:

maximum order = 5
probe = Z116098_0395
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 13C
o1p (from lock) = 7.24 ppm
optimisation parameters = ls
linewidth 1H = 0.10 Hz
envelope shape / strictness = 2.00 / 200

Results:

tune lock level increase = 1.2

initial B0 stdDev = 4.53 Hz
sample size = 2.81 cm, position = 0.04 cm
final B0 stdDev = 0.31 Hz > improvement = 14.8
envelope width = 0.67 Hz

Performing tune after shimming
tune lock level increase = 1.0

shim changes: 
Z	+63
Z2	-56
Z3	+99
Z4	-420
Z5	+175
X	-44
Y	+126

duration = 3 min 31 sec
completed successfully
finished Tue Oct 31 17:35:09 2023


========================================================================